Keywords: Amplatzer cardiac plug; Left atrial appendage occlusion; Device embolization
The following are the defined steps for ACP plug preparation:
a) Immerse device & hub end of loader in sterile saline to remove air.
b) Actively manipulate device in saline by hand to eliminate air bubbles on device.
c) Pull loading cable vice until lobe is fully retracted within loader, & stop before disc is recaptured
d) Insert distal end of delivery cable through haemostasis valve.
e) Connect delivery cable to exposed proximal end screw of device.
f) Grasp hub & rotate delivery cable clockwise until device is fully threaded on delivery cable.
g) Rotate delivery cable counter clockwise 1/8 of one turn to ensure that cable is not over tightened.
Numerous complications can be avoided if these steps are the done carefully and with utmost attention to every minute detail. In our case, we assume that excessive reverse rotation of the cable might be responsible for the device being loosely connected to the cable which got easily disconnected by the tug.